Submit your papersSubmit Now
For Enquiries: [email protected]
IIARD LogoIIARD

Fatigue Analysis of Varying Femoral Head Diameters in Total Hip Implants Using Finite Element Method

Komonibo, Ernest Bolouebi, and Brisibe Endurance, ., Agriculture, Nembe Bassambiri, Bayelsa State, Nigeria.

Abstract

Mechanical endurance is recognized as a critical design criterion for ensuring the optimal operational performance of hip implants. Consequently, the primary objective of this research is to assess the fatigue behaviour of hip implants across a range of femoral head diameters. Specifically, the investigation aimed to elucidate how variations in prosthetic femoral head dimensions affect the cyclic durability of total hip replacement prostheses. To achieve this, seven distinct hip implant models, fabricated from the Ti-6Al-4V alloy and featuring femoral head diameters ranging from 28mm to 52mm, were developed using SolidWorks Premium 2025. These models were then subjected to finite element analysis via ANSYS 2022 R-2 to characterize their respective fatigue responses. The FEA results demonstrated an inverse correlation between the fatigue life (expressed as the number of cycles) and the femoral head diameter. Specifically, cyclic endurance diminished proportionally with increasing femoral head dimensions. The peak cyclic longevity, recorded at 1.120e7 cycles, was associated with the 28mm femoral head diameter, while the lowest cyclic longevity of 5.016e6 cycles corresponded to the 52mm femoral head diameter. Thus, the findings of this research comprehensively illustrate the impact of varying femoral head diameters on the fatigue life of hip implants.

Keywords

Femoral Head DiameterHip ImplantFinite Element AnalysisAnsysFatigue life.

References

Kaya, F., & İnce, G. (2024). Fatigue Analysis on a Newly Designed Hip Implant with Finite Element Method. International Journal of Engineering and Innovative Research, 6(3), 162-178. Marsh, M., & Newman, S. (2021). Trends and developments in hip and knee arthroplasty technology. Journal of rehabilitation and assistive technologies engineering, 8, 2055668320952043. Eachempati, K. K., Dannana, C. S., Apsingi, S., Ponnala, V. K., Boyapati, G., & Parameswaran, A. (2020). Trunnion fracture of femoral prosthesis following a large metal-on-metal uncemented total hip arthroplasty: a case report. Arthroplasty, 2(1), 32. Grupp, T. M., Weik, T., Bloemer, W., & Knaebel, H. P. (2010). Modular titanium alloy neck adapter failures in hip replacement-failure mode analysis and influence of implant material. BMC musculoskeletal disorders, 11(1), 3. Janssen, D., Aquarius, R., Stolk, J., & Verdonschot, N. (2005). Finite-element analysis of failure of the Capital Hip designs. The Journal of Bone & Joint Surgery British Volume, 87(11), 1561-1567. Łapaj, Ł., Woźniak, W., Wiśniewski, T., Rozwalka, J., Paczesny, Ł., Zabrzyński, J., ... & Kruczyński, J. (2019). Breakage of metal hip arthroplasty components: Retrieval and structural analysis. Bio-Medical Materials and Engineering, 30(3), 297-308. Mierzejewska, Ż. A. (2015). Case study and failure analysis of a total hip stem fracture. Advances in materials science, 15(2), 5-13. Joshi, T., Sharma, R., Mittal, V. K., Gupta, V., & Krishan, G. (2022). Dynamic fatigue behavior of hip joint under patient specific loadings. International Journal of Automotive and Mechanical Engineering, 19(3), 10014-10027. Pekedis, M., & Yildiz, H. (2011). Comparison of fatigue behaviour of eight different hip stems: a numerical and experimental study. Journal of Biomedical Science and Engineering, 4(10), 643-650. Zameer, S., & Haneef, M. (2015). Fatigue life estimation of artificial hip joint model using finite element method. Materials Today: Proceedings, 2(4-5), 2137-2145. Milovanović, A., Sedmak, A., Grbović, A., Mijatović, T., & Čolić, K. (2020). Design aspects of hip implant made of Ti-6Al-4V extra low interstitials alloy. Procedia Structural Integrity, 26, 299-305. Bhawe, A. K., Shah, K. M., Somani, S., Shenoy B, S., Bhat N, S., Zuber, M., & KN, C. (2022). Static structural analysis of the effect of change in femoral head sizes used in Total Hip Arthroplasty using finite element method. Cogent Engineering, 9(1), 1-13. Dimple, D., Shruti, M., Badgayan, N. D., & Sahu, S. K. (2021). Finite element analysis of hdpe- based hybrid nanocomposite for potential use as liner material for total hip prosthesis. In Advances in Engineering Design: Select Proceedings of FLAME 2020, 305-313. Liu, S., & Shin, Y. C. (2019). Additive manufacturing of Ti6Al4V alloy: A review. Materials & Design, 164, 107552. Sachin, G., & Yogesh, j. (2022). Biomaterials in Hip Joint Replacement. Journal of Engineering Research and Sciences1, 1(8), 01–11.